0025 00
TM 1-6625-724-13&P
0025 00-14
NOTE
Errors are returned from a variety of sources and are
meant to diagnose system problems. Sources of errors
can be equipment setup errors, software setup errors,
internal hardware failures, or internal software failures.
Each error type contains an error code number, a text
description of the error, and proposes corrective action.
NOTE
Operating System (OS9) error codes are not listed
below. If an OS9 error code is displayed, perform
actions displayed and contact factory for support.
Table 1. Error Reported on CADU Error Screen
Code #
Message
4097:
Vibration FIFO Test Failure ACTION: (1) Power the DAU off and on (2) Wait 20 seconds,
make a measurement
4098:
Unstable Ground Reference ACTION: (1) Power the DAU off and on (2) Wait 20 seconds,
make a measurement
4099:
Unstable Maximum Reference ACTION: (1) Power the DAU off and on (2) Wait 20 seconds,
make a measurement
4100:
Unstable Minimum Reference ACTION: (1) Power the DAU off and on (2) Wait 20 seconds,
make a measurement
4101:
Ground Ref Offset Error ACTION: (1) Power the DAU off and on (2) Wait 20 seconds, make
a measurement
4102:
Maximum Ref Gain Error ACTION: (1) Power the DAU off and on (2) Wait 20 seconds, make
a measurement
4103:
Minimum Ref Gain Error ACTION: (1) Power the DAU off and on (2) Wait 20 seconds, make
a measurement
4104:
A/D Consistency Error ACTION: (1) Power the DAU off and on (2) Wait 20 seconds, make a
measurement
The A/D consistency check failed on power up. Power the DAU off and on, and then try to make
a measurement. If the failure persists, send unit to TMDE for repair. Suspect failure of the DAU
Acquisition board.
4105:
A/D Dynamic Performance Error ACTION: (1) Power the DAU off and on (2) Wait 20
seconds, make a measurement
The A/D dynamic performance check failed on power up. Power the DAU off and on, and then
try to make a measurement. If the failure persists, send unit to TMDE for repair. Suspect failure
of the DAU Acquisition board.