TM 1-6625-724-13&P0025 000025 00-17Table 1. Error Reported on CADU Error Screen (cont.)Code #Message4127:DAC Filter Ripple Error ACTION: (1) Power the DAU off and on (2) Wait 20 seconds, makea measurement (3) If error persists, possible hardware failureThe internal DAC filter test has failed on calibration. Power the DAU off and on, and then try tomake a measurement. If the failure persists, send unit to TMDE for repair.4128:Filter 1 Gain Tolerance ACTION: (1) Power the DAU off and on (2) Wait 20 seconds, make ameasurement (3) If error persists, possible hardware failureThe filter 1 gain tolerance is larger than 2% and has failed its calibration test on power up.Power the DAU off and on, and then try to make a measurement. If the failure persists, send unitto TMDE for repair. Suspect failure of DAU Acquisition board.4129:Filter 2 Gain Tolerance ACTION: (1) Power the DAU off and on (2) Wait 20 seconds, make ameasurement (3) If error persists, possible hardware failureThe filter 2 gain tolerance is larger than 2% and has failed its calibration test on power up.Power the DAU off and on, and then try to make a measurement. If the failure persists, send unitto TMDE for repair. Suspect failure of DAU Acquisition board.4130:Filter 3 Gain Tolerance ACTION: (1) Power the DAU off and on (2) Wait 20 seconds, make ameasurement (3) If error persists, possible hardware failureThe filter 3 gain tolerance is larger than 2% and has failed its calibration test on power up.Power the DAU off and on, and then try to make a measurement. If the failure persists, send unitto TMDE for repair. Suspect failure of DAU Acquisition board.4131:Filter 4 Gain Tolerance ACTION: (1) Power the DAU off and on (2) Wait 20 seconds, make ameasurement (3) If error persists, possible hardware failureThe filter 4 gain tolerance is larger than 2% and has failed its calibration test on power up.Power the DAU off and on, and then try to make a measurement. If the failure persists, send unitto TMDE for repair. Suspect failure of DAU Acquisition board.4132:Filter 5 Gain Tolerance ACTION: (1) Power the DAU off and on (2) Wait 20 seconds, make ameasurement (3) If error persists, possible hardware failureThe filter 5 gain tolerance is larger than 2% and has failed its calibration test on power up.Power the DAU off and on, and then try to make a measurement. If the failure persists, send unitto TMDE for repair. Suspect failure of DAU Acquisition board.4133:Filter 6 Gain Tolerance ACTION: (1) Power the DAU off and on (2) Wait 20 seconds, make ameasurement (3) If error persists, possible hardware failureThe filter 6 gain tolerance is larger than 2% and has failed its calibration test on power up.Power the DAU off and on, and then try to make a measurement. If the failure persists, send unitto TMDE for repair. Suspect failure of DAU Acquisition board.4144:Multiple Measurement Failure: view error log ACTION: (1) Power the DAU off and on (2)Wait 20 seconds, make a measurement4145:Accelerometer Failure on Input 1 ACTION: (1) Check cable (2) Check accelerometer (3)Verify SETUP (also go to Setup/Status in Manager Menu)4146:Accelerometer Failure on Input 2 ACTION: (1) Check cable (2) Check accelerometer (3)Verify SETUP (also go to Setup/Status In Manager Menu)4147:Accelerometer Failure on Input 3 ACTION: (1) Check cable (2) Check accelerometer (3)Verify SETUP (also go to Setup/Status In Manager Menu)
Integrated Publishing, Inc. - A (SDVOSB) Service Disabled Veteran Owned Small Business