TM 1-6625-724-13&P0037 000037 00-1CADU LCD ASSEMBLYREMOVAL AND INSTALLATIONCAUTIONThis assembly contains parts sensitive to damageby Electrostatic Discharge (ESD).Use precautionary ESD procedures when touching,removing or inserting. Use static-free material towrap the assembly for shipment or storage.IndexRemoval............................................................................................................................................................................ 1Installation......................................................................................................................................................................... 2InitialSetup:Workbench with ESD capabilityPersonnelRequired:TMDE Support SpecialistToolsandEquipment:Tool Kit, ElectriciansParts/Materials:LCD Assy PN 29324501, Threadlock adhesive, Permabond 910, Lens cleanerEquipmentCondition:CADU power turned off. Lay unit with keypad down on support strips toprotect facing.CAUTIONThis assembly contains parts sensitive to damageby Electrostatic Discharge (ESD).Use precautionary ESD procedures when touching,removing or inserting. Use static-free material towrap the assembly for shipment or storage.a. REMOVAL: 1. Remove nine screws (1), nine lock washers (2), and nine flat washers (3) from bottom plate (4). 2. Remove bottom plate (4) from top case (5). 3. Disconnect battery cable connector (6) from J6 on the PWA.