0025 00-24
Table 1. Error Reported on CADU Error Screen (cont.)
Code #
Message
4249:
Accelerometer Saturation on Accelerometer Channel 11 ACTION: (1) Verify Setup (Setup
Status In Manager Menu) (2) Change accelerometer Type (3) Relocate accelerometer
4250:
Accelerometer Saturation on Accelerometer Channel 12 ACTION: (1) Verify Setup (Setup
Status In Manager Menu) (2) Change accelerometer Type (3) Relocate accelerometer
4251:
Accelerometer Saturation on Accelerometer Channel 13 ACTION: (1) Verify Setup (Setup
Status In Manager Menu) (2) Change accelerometer Type (3) Relocate accelerometer
4252:
Accelerometer Saturation on Accelerometer Channel 14 ACTION: (1) Verify Setup (Setup
Status In Manager Menu) (2) Change accelerometer Type (3) Relocate accelerometer
8193:
Reception Not Acknowledged ACTION: (1) Retry measurement (2) Check CADU to DAU
cable (3) Check that DAU's power switch Is on (4) Reboot CADU
8194:
Spkt missed a Recvack Event ACTION: (1) Retry measurement (2) Check CADU to DAU
cable (3) Check that DAU's power switch is on (4) Reboot CADU
8195:
Spkt Missed a Xmitack Event ACTION: (1) Retry measurement (2) Check CADU to DAU
cable (3) Check that DAU's power switch is on (4) Reboot CADU
8196:
Illegal Communication Channel Number ACTION: (1) Reboot CADU (2) Repeat key
sequence (3) If error reoccurs, Please report error to RADS-AT manufacturer (instead of
reporting error to manufacturer, report error to AVSCOM).
8197:
Data Transmission Error ACTION: (1) Retry measurement (2) Check CADU to DAU cable
(3) Check that DAU's power switch Is on (4) Reboot CADU
8198:
Data Reception Error ACTION: (1) Retry measurement (2) Check CADU to DAU cable (3)
Check that DAU's power switch Is on (4) Reboot CADU
8199:
Communications Failure ACTION: (1) Retry measurement (2) Check CADU to DAU cable
(3) Check that DAU's power switch is on (4) Reboot CADU
16385:
Measurement Aborted
The measurement has been aborted due to user initiated abort or other failure. This message is
usually a reported after another error message.
16386:
Too Many Samples Out-of-Range ACTION: (1) Power the DAU off and on (2) Wait 20
seconds, make a measurement
The DAU has detected too many samples that saturate the internal sampling circuit. Repeat the
measurement. This error is caused by amplitude ranges changing drastically during acquisition.
May be caused by extreme turbulence.
16387:
Sensor Fault on Internal Channel 1 ACTION: (1) Check cable (2) Check accelerometer -
find channel associated with position number above, In script file
16388:
Sensor Fault on Internal Channel 2 ACTION: (1) Check cable (2) Check accelerometer -
find channel associated with position number above, In script file
16389:
Sensor Fault on Internal Channel 3 ACTION: (1) Check cable (2) Check accelerometer -
find channel associated with position number above, In script file